氮氧化物NO+NO2氣體檢測管175SH,氮氧化物NO+NO2氣體檢測管175SH NITROGEN OXIDES 100-2500ppm,日本光明理化工業(yè)株式會社從1947年開始,在氣體分析領域開拓了六十余年 , 形成了馳名于世的 ” 北川 ” 式氣體檢測管。氣體快速分析法 ” 北川 ” 的歷史就是日本檢測管的歷史 。目前是世界上*大的檢測管生產(chǎn)廠家,也是日本**家通過美國SEI的認證,歐盟的認證的檢測管廠家。公司不僅有ISO9001:2000的認證,還獲得了環(huán)境管理的ISO14001的認證。北川式手動氣體采集器AP-20 備有采氣完成指示計 , 可以清楚的判明采氣是否完成 . • 有專門截留玻璃的截留槽 , 以防止檢測管割尖時玻璃片飛濺 . • 有采氣容積為 50 毫升和 100 毫升的兩段式制動卡 . • 采氣器極為小巧 , 并有便于攜帶的皮具包 。
Tube No.175SH NITROGEN OXIDES
1. PERFORMANCE
1) Measuring range :100-2,500 ppm
Number of pump strokes 1(100ml)
2) Sampling time :1 minute/1 pump stroke
3) Detectable limit :10 ppm
4) Shelf life :1 year
5) Operating temperature :5~45℃
6) Reading :Direct reading from the scale calibrated by 1 pump stroke
7) Colour change :White→Green
2. RELATIVE STANDARD DEVIATION
RSD-low:10% RSD-mid.:10% RSD-high:10%
3. CHEMICAL REACTION
NO ; By reacting with an Oxidizer, NO2 is produced.
NO+CrO3+H2SO4→NO2
NO2+(C6H5)2NH→(C6H5)2NNO
NO2 ; By reacting with Diphenylamine ,N- Nitroso-diphylamine is produced.
NO2+(C6H5)2NH→(C6H5)2NNO
4. CALIBRATION OF THE TUBE
NO ; STANDARD GAS CYLINDER METHOD
NO2 ; PERMEATION TUBE METHOD
5. INTERFERENCE AND CROSS SENSITIVITY
100
200
300
400
600
800
1000
1500
2000
2500
ppm
100ml
GAS To pump
PRETREAT REAGENT
(Reddish brown)
Substance Interference ppm Coexistence
Hydrogen chloride Tchhaen bgoetdto tmo Dofa drkis bclouleo.ured layer is 500 Higher readings are given.
Sulphur dioxide The accurecy of reading is not
affected.
(NOTE)
When the concentration of Nitrogen oxides is high(over 2,000 ppm)in the measuring range, a green ring may
occur in the discoloured layer or double-stained layer may occur. The total stain length should be read, even if
the stained layer gets multi-colour discolouration.
1. PERFORMANCE
1) Measuring range :100-2,500 ppm
Number of pump strokes 1(100ml)
2) Sampling time :1 minute/1 pump stroke
3) Detectable limit :10 ppm
4) Shelf life :1 year
5) Operating temperature :5~45℃
6) Reading :Direct reading from the scale calibrated by 1 pump stroke
7) Colour change :White→Green
2. RELATIVE STANDARD DEVIATION
RSD-low:10% RSD-mid.:10% RSD-high:10%
3. CHEMICAL REACTION
NO ; By reacting with an Oxidizer, NO2 is produced.
NO+CrO3+H2SO4→NO2
NO2+(C6H5)2NH→(C6H5)2NNO
NO2 ; By reacting with Diphenylamine ,N- Nitroso-diphylamine is produced.
NO2+(C6H5)2NH→(C6H5)2NNO
4. CALIBRATION OF THE TUBE
NO ; STANDARD GAS CYLINDER METHOD
NO2 ; PERMEATION TUBE METHOD
5. INTERFERENCE AND CROSS SENSITIVITY
100
200
300
400
600
800
1000
1500
2000
2500
ppm
100ml
GAS To pump
PRETREAT REAGENT
(Reddish brown)
Substance Interference ppm Coexistence
Hydrogen chloride Tchhaen bgoetdto tmo Dofa drkis bclouleo.ured layer is 500 Higher readings are given.
Sulphur dioxide The accurecy of reading is not
affected.
(NOTE)
When the concentration of Nitrogen oxides is high(over 2,000 ppm)in the measuring range, a green ring may
occur in the discoloured layer or double-stained layer may occur. The total stain length should be read, even if
the stained layer gets multi-colour discolouration.